TB 9-6625-2071-35
Calibrator output
Test instrument
settings
indications (V)
(V dc)
Min
Max
Adjustments
9.5
9.5033
9.4967
b(2)
-9.5
-9.5033
-9.4967
b(3)
95
+94.967
+95.033
b(4)
950
+949.13
+950.87
b(5)
b. Adjustments
(1) Adjust A6R29 (DVM zero adjustment) through access hole on front panel for a
TI indication between -0.0001 and +0.0001 V (R).
(2) Adjust A7R13 (fig. 1) for a TI indication between +9.4999 and +9.5001 V (R).
(3) Adjust A7R6 (fig. 1) for a TI indication between -9.4999 and -9.5001 V (R).
(4) Adjust A6R29 (fig. 1) for a TI indication between +94.999 and +95.001 (R).
(5) Adjust A6R15 (fig. 1) for a TI indication between +949.99 and +950.01 (R).
13. Remote Trigger Level (Option 041, AN/USM-459, and
5328AF096)
a. Performance Check
NOTE
Perform remote trigger level only if TI is used in an automated
configuration.
NOTE
Remote trigger level can only be performed in an automated
mode. The TI must be connected to a controller with a properly
configured GPIB IEEE interface card.
NOTE
If you are using a controller and/or software other than the
CALSETS fielded controller and ICE software, you will have to
execute the TI command strings as required.