TB 9-6625-2071-35
(16) Execute TI command string PF4G6S1S3A379-200*B37-200*R.
(17) Set function generator Offset to -2 V dc.
indicates -2.000 V dc 0.002 V dc.
(fig. 1) for a 50
percent duty
cycle on
oscilloscope display (R).
(20) Move oscilloscope Vertical 1 connection from TI CHANNEL A MARKER
OUTPUT to TI CHANNEL B MARKER OUTPUT.
(fig. 1) for a 50
percent duty
cycle on
oscilloscope display (R).
(22) Set function generator Offset to 0 V dc.
a. Deenergize and disconnect all equipment and reinstall TI protective cover.
b. Annotate and affix DA label/form in accordance with TB 750-25.