Table 8. Volts/Div Accuracy CH 2
Table 10. CH 1 and CH 2 High Frequency Compensation Adjustments

Calibration Procedure For Oscilloscope, Types 453, 453A, 454, 454Mod163D, And 468 Page Navigation
  18    19    20    21    22  23  24    25    26    27    28  
img
. . . . . . . . . . . . . . .
TB 9-6625-2083-24
NOTE
If standardizer does not have sufficient range to obtain
optimum square wave, adjust C9 (fig. 6) to midrange and
readjust standardizer (R).
Figure 7. Bottom and left side view - adjustment locations.
(4) Repeat technique of (3) above for VOLTS/DIV switch settings and adjustments
listed in table 9, but do not readjust standardizer.
(5) Set oscilloscope calibrator for a CHAN 2, EDGE mode output of 100 mV at 1 kHz.
(6) Set TI MODE switch to CH 2 and repeat technique of (3) and (4) above, using
CH 2 controls and adjustments listed in table 9.
Table 9. Volts/Div Switch Compensation
Test instrument
Adjustments (R)
VOLTS/DIV
Square corner
Flat top
switch
CH 1
CH 2
CH 1
CH 2
settings
0.1
---
---
C9
C109
(fig. 6)
(fig. 6)
0.2
C5D
C105D
C5A
C105A
(fig. 7)
(fig. 7)
(fig. 7)
(fig. 7)
21
C6E
C106E
C6A
C106A
(fig. 7)
(fig. 7)
(fig. 7)
(fig. 7)
1Remove
termination.
25