TB 9-6625-2083-24
Test instrument
A AND B TIME/DIV AND
MARKER mode
Err display
DELAY TIME switch
output
indication limits (%)
settings
0.1 s
0.1 s
3
0.2 s
0.2 s
3
0.5 s
0.5 s
1
1
3
s
s
2
2
1
1
3
s
s
2
2
3
s
s
5
5
3
s
s
10
10
3
s
s
20
20
3
s
s
50
50
3
0.1 ms
0.1 ms
3
0.2 ms
0.2 ms
3
0.5 ms
0.5 ms
3
1 ms
1 ms
3
2 ms
2 ms
3
5 ms
5 ms
3
3
3
10 ms
10 ms
3
20 ms
20 ms
3
50 ms
50 ms
3
0.1 s
0.1 s
3
0.2 s
0.2 s
3
0.5 s
0.5 s
3
1 s4
1 s4
3
4
4
2 s
2 s
3
4
4
5 s
5 s
3
1Adjust
C740A (fig. 8) for A TIME/DIV and C940A (fig. 8) for B TIME/DIV AND DELAY TIME (R).
2Adjust
C740C (fig. 8) for A TIME/DIV and C940 C (fig. 8) for B TIME/DIV AND DELAY TIME (R).
3Set A SWEEP MODE switch to NORM TRIG.
4A SWEEP only.
b. Adjustments. Perform adjustments listed in table 11 footnotes.
a. Performance Check
(1) Position switches as listed in (a) through (c) below:
(a) CH 1 VOLTS/DIV to 50 mV.
(b) A AND B TIME/DIV AND DELAY TIME to .05 s.
(c) HORIZ DISPLAY to A.
(2) Set oscilloscope calibrator for a CHAN 1, MARKER mode output of 5 ns.
(3) Position MAG switch to X10.
(4) Position second marker at second graticule line so that sweep starts at first
graticule.
(5) If display does not have optimum linearity (even spacing) over center 8 divisions
of crt, perform b (1) below.