TB 9-6625-2083-24
b. Adjustments
(1) Rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton for
an Err display of 0.0%. Adjust R1036 (fig. 8) until 10th marker aligns with ninth vertical
graticule line (R).
(2) Rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton for
an Err display of 0.0%. Adjust R1047 (fig. 8) for 1 marker per division (R).
(3) Adjust R1053 (fig. 8) for no shift of display as MAG switch is set from X10
to OFF (R).
a. Performance Check
(1) Position controls as listed in (a) through (d) below:
DELAY TIME MULTIPLIER control to 0.10.
(a)
B TIME/DIV AND DELAY TIME switch to 1 ms.
(b)
B SWEEP MODE switch to TRIGGERABLE AFTER DELAY TIME.
(c)
HORIZ DISPLAY switch to B (DELAYED SWEEP).
(d)
(2) Set oscilloscope calibrator for a MARKER mode output of 1 ms.
(3) Adjust B LEVEL control for a stable display.
(4) Adjust horizontal POSITION control and rotate oscilloscope calibrator knob
located below EDIT FIELD pushbutton to align 1 marker per division on the middle 8
divisions. If oscilloscope calibrator Err display readout does not indicate within 3%,
perform b below.
b. Adjustments. Rotate oscilloscope calibrator knob located below EDIT FIELD
pushbutton for an Err display of 0.0%, and adjust R943 (fig. 8) for 1 marker per division (R).
a. Performance Check
(1) Set HORIZ DISPLAY switch to A.
(2) Set A AND B TIME/DIV AND DELAY TIME switches to .05 s.
(3) Set oscilloscope calibrator for a CHAN 1, MARKER mode output of 50 ns.
(4) Adjust horizontal POSITION control and rotate oscilloscope calibrator knob
located below EDIT FIELD pushbutton to align 1 time marker per division on the middle 8
divisions. If oscilloscope calibrator Err display readout does not indicate within limits
specified in first row of table 11, perform b below.
(5) Repeat technique of (4) above, using settings and indications listed in table 11.
If oscilloscope calibrator Err display readout does not indicate within limits specified in
(6) Set HORIZ DISPLAY switch to B (DELAYED SWEEP).
(7) Repeat technique of (3) through (5) above for B DELAYED SWEEP.