Equipment Setup
Table 14. Vertical Gain CH 1

Calibration Procedure For Oscilloscope, Types 453, 453A, 454, 454Mod163D, And 468 Page Navigation
  28    29    30    31    32  33  34    35    36    37    38  
img
. . . . .
TB 9-6625-2083-24
Figure 9. Oscilloscope left side view.
36. Vertical Gain
a. Performance Check
(1) Set CH 1 and CH 2 AC-GND-DC switches to DC.
(2) Connect oscilloscope calibrator SOURCE/MEASURE CHAN 1 to TI CH1 input
and SOURCE/MEASURE CHAN 2 to TI CH2 input through 50
feedthrough
terminations.
(3) Set oscilloscope calibrator for a CHAN 2, VOLTAGE mode output of 30 mV and
frequency of 1 kHz.
(4) Rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton for 6
divisions of TI display.  If oscilloscope calibrator Err display readout indication is not
within limits specified in first row of table 13, perform b (1) below.
(5) Set VOLTS/DIV and oscilloscope calibrator VOLTAGE to settings listed in
table 13 and at each setting, rotate oscilloscope calibrator knob located below EDIT FIELD
35