Figure 10. Storage display board
Storage Acquisition Step Response

Calibration Procedure For Oscilloscope, Types 453, 453A, 454, 454Mod163D, And 468 Page Navigation
  33    34    35    36    37  38  39    40    41    42    43  
img
. . . . . .
TB 9-6625-2083-24
Table 17. Low Frequency Compensation
Test instrument
Adjustments
Oscilloscope calibrator
TIME/DIV
EDGE output
switch settings
CH 1
CH 2
R1651
R3711
0.2 ms
1 kHz
s
R1721
R3731
10 kHz
20
s
R3211
R3211
100 kHz
2
R3191
R3191
1Adjust
R165, R172, R371, and R373 (fig. 9), and R319 and R321 (fig. 11) for best compromise between CH 1 and CH
Figure 11. Vertical output board.
39. CH 1 and CH 2 VOLTS/DIV Compensation
a. Performance Check
(1) Position controls as listed in (a) through (c) below:
(a) VERT MODE CH 1 pushbutton pressed and CH 2 pushbutton released.
(b) 20 MHz BW LIMIT pushbutton pressed to 20 MHz (in).
(c) A AND B TIME/DIV AND DELAY TIME switches to 0.2 ms.
(2) Connect oscilloscope calibrator SOURCE/MEASURE CHAN 1 to CH 1 input
and SOURCE/MEASURE CHAN 2 to CH 2 input through 50  feedthrough terminations,
attenuators, and standardizers.
(3) Set oscilloscope calibrator for a CHAN 1, EDGE mode output of 25 mV at 1
kHz, and rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton for 5
divisions of display.
(4) Adjust standardizer for best square wave display. Display will be optimum
square wave without rounding or overshoot in excess of one-half minor division.
40