Table 19. A and B Timing
Figure 13. Timing board.

Calibration Procedure For Oscilloscope, Types 453, 453A, 454, 454Mod163D, And 468 Page Navigation
  38    39    40    41    42  43  44    45    46    47  
img
. .
TB 9-6625-2083-24
(4) Rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton for
an Err display of 0.0%. Adjust R571 X10 GAIN (fig. 12) for 1 time marker per division (R).
Figure 12. Interface board.
45