TB 9-6625-2083-24
(5) Repeat technique of (3) and (4) above, except do not readjust standardizer, for
VOLTS/DIV switch settings listed in table 18. If TI does not display optimum square
wave, perform appropriate adjustments listed in table 18.
(6) Press VERT MODE CH 2 and release CH 1 pushbuttons.
(7) Set oscilloscope calibrator for a CHAN 2, EDGE mode output of 25 mV at 1 kHz
and repeat technique of (3) through (5) above for CH 2 switch settings, except do not
readjust standardizer.
b. Adjustments. No further adjustments can be made.
Oscilloscope calibrator
Test instrument
EDGE mode
VOLTS/DIV
Adjustments
amplitude
switch setting
Flat top
Corner
1
25 mV
5 mV
C13010
-----
50 mV
10 mV
C37
C36
100 mV
20 mV
C35
C34
250 mV
50 mV
C33
C32
2
0.5 V
0.1 V
-----
-----
1V
0.2 V
-----
-----
2.5 V
0.5 V
C31
C30
1CH
2 only.
40. Storage Acquisition Step Response
a. Performance Check
(1) Position controls as listed in (a) through (e) below:
VOLTS/DIV switches to 5m.
(a)
A AND B TIME/DIV AND DELAY TIME switches to .1 s.
(b)
STORAGE MODE NORM pushbutton pressed to ON.
(c)
DISPLAY RESPONSE pushbutton released to PULSE.
(d)
STORAGE WINDOW pushbutton released to POST TRIG.
(e)
(2) Connect oscilloscope calibrator SOURCE/MEASURE CHAN 2 to CH 2 input,
through 50 feedthrough termination.
(3) Set oscilloscope calibrator for a CHAN 2, EDGE mode output of 25 mV at 100
KHz and rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton for 5
divisions of display on TI. Press ENVELOPE STORAGE MODE pushbutton and then
press SAVE pushbutton.
(4) After envelope acquisition cycle completes, check that save waveform has less
than 1 minor division of overshoot on leading edge of waveform.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Position controls as listed in (a) through (h) below: