Table 16. Volts Cursor Readout Accuracy CH2
Figure 11. Vertical output board.

Calibration Procedure For Oscilloscope, Types 453, 453A, 454, 454Mod163D, And 468 Page Navigation
  32    33    34    35    36  37  38    39    40    41    42  
img
. . . . . . .
TB 9-6625-2083-24
Figure 10. Storage display board
38. Low Frequency Compensation
a. Performance Check
(1) Position controls as listed in (a) through (e) below:
(a) VOLTS/DIV switches to 5 mV.
(b) A AND B TIME/DIV AND DELAY TIME switches to 0.2 ms.
(c) VERT MODE pushbutton to CH 1 and CH 2 pushbutton released.
(d) AC-GND-DC switches to DC.
(e) NON STORE pushbutton pressed.
(2) Set oscilloscope calibrator for a CHAN 1, EDGE mode output of 25 mV at 1 kHz
and rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton for a 5
division display on TI. If square wave displayed has aberrations of 1 minor division or
more, perform appropriate adjustments listed in table 17.
(3) Set oscilloscope calibrator CHAN 1, EDGE mode output and TI TIME/DIV
switch to settings listed in table 17 and repeat technique of (2) above for switch settings
listed in table 17.
(4) Set oscilloscope calibrator for a CHAN 2, EDGE mode output. Press TI VERT
MODE pushbutton to CH 2, release CH 1 pushbutton, and repeat technique of (2) above
for CH 2 switch settings and adjustments listed in table 17.
b. Adjustments. No further adjustments can be made.
39