TB 9-6625-2094-24
(1) Connect measuring receiver sensor module to TI RF OUTPUT.
(2) Configure measuring receiver to measure RF power in LOG mode.
(3) Position controls as listed in (a) through (e) below:
(a)
Set FREQUENCY switches to indicate 050.000 MHz.
(b)
MODULATION MODE switch to CW.
(c)
OUTPUT VERNIER control for a -1 dB indication on OUTPUT meter.
(d)
OUTPUT step attenuator control to +10 dBm.
(e)
(4) If measuring receiver does not indicate within limits specified in first row of
(5) Repeat technique of (3) (d) and (e) above for TI settings and indications listed in
table 4. If measuring receiver does not indicate within specified limits, perform b below.
Test instrument
Power meter (dBm)
OUTPUT step attenuator
OUTPUT meter
settings (dBm)
settings
Min
Max
+10
-1
+8.5
+9.5
0
+3
+2.5
+3.5
0
0
-0.5
+0.5
0
-3
-2.5
-3.5
0
-5
-4.5
-5.5
0
-7
-6.5
-7.5
-10
+3
-6.5
-7.5
b. Adjustments
(1) Temporarily disconnect measuring receiver sensor module from TI.
(2) Set OUTPUT VERNIER control fully ccw and adjust meter board POT B (fig. 2)
until meter needle bisects dot at left end of meter scale (R).
(3) Adjust OUTPUT VERNIER control fully cw and adjust meter board POT A (fig.
2) for a +3-dBm OUTPUT meter reading (R).
(4) Set OUTPUT step attenuator to +10 dBm.
(5) Connect measuring receiver sensor module to TI RF OUTPUT.
(6) Alternately adjust OUTPUT VERNIER control for a 0 and -7 dBm reading on
TI OUTPUT meter while adjusting meter board POT F (fig. 2) for +10 dBm and POT E (fig.
2) for +3 dBm indications on the measuring receiver. Repeat the above until no further
adjustments are required (R).
10