TB 9-6625-2097-35
Table 1. Calibration Description
Test instrument
Performance
parameters
specifications
Recorders, Ink Writing, Model 2200S (Series)
Input signal
Range: 0 to 5 V at dc to 50 Hz
Frequency response
Range: 100 div at 100 mm amplitude, dc to 30 Hz
1 div
Accuracy:
Range: 50 div at 50 mm amplitude, dc to 50 Hz
1 div
Accuracy:
Overshoot
Range: 100 mm amplitude
Accuracy:
Less than 8 ms
Range: 50 mm amplitude
Accuracy:
Less than 5 ms overshoot
less than 1% between 10 to 90% FS
Input power
100, 115, 200, and 230 V, 10%, 50 to 400 Hz
RMS Converter, Model 13-4618-10
Converter output
Range: 0 to 5 V dc linear
0.2% of reading when reference to input
Accuracy:
voltage with accuracy of 0.05%
Converter input
Range: X1 V .05 to 10 V
X100 V 5 to 1000 V
Frequency Deviation Converter, Model 13-4618-00
Frequency Deviation
Range: .5, 1, 2.5, and 5 Hz at 50 or 60 Hz center frequency
5, 10, 25, and 50 Hz at 400 Hz center frequency
Output voltage:
2.5 V dc FS proportional to frequency
deviation at each range switch setting
0.3% when input frequency is 0.075%
Accuracy:
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2
identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286. Alternate items may be used by the calibrating activity when
the equipment listed in table 2 is not available. The items selected must be verified to
perform satisfactorily prior to use and must bear evidence of current calibration. The
equipment must meet or exceed the minimum use specifications listed in table 2. The
accuracies listed in table 2 provide a four-to-one ratio between the standard and TI.
5. Accessories Required. The accessories listed in table 5 are issued as indicated in
paragraph 4 above and
are used in this
calibration procedure. When necessary, these items
may be substituted by equivalent items, unless specifically prohibited.
3