TB 9-6625-2097-35
Paragraph
Page
Equipment setup ..............................................
Converter output ..............................................
Final procedure ................................................
SECTION
CALIBRATION PROCESS FOR
FREQUENCY DEVIATION CONVERTER,
MODEL 13-4618-00
Preliminary instructions ..................................
Equipment setup ..............................................
Span gain..........................................................
Final procedure ................................................
SECTION I
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Recorders, Ink Writing, Gould, Model 2200S (Series); with RMS Converter,
Gould, Model 13-4618-10; and Frequency Deviation Converter, Gould, Model 13-4618-00.
The manufacturer's manuals were used as the prime data sources in compiling these
instructions. The equipment being calibrated will be referred to as the TI (test instrument)
throughout this bulletin.
a. Model Variations. None.
b. Time and Technique. The time required for this calibration is approximately 3
hours, using the dc and low frequency technique.
2. DA Form 2416 (Calibration Data Card)
a. Forms, records, and
reports required for calibration personnel at all
levels are
prescribed by TB 750-25. DA Form 2416 must be annotated in accordance with TB 750-25
for each calibration performed.
b. Adjustments to be reported on DA Form 2416 are designated (R) at the end of the
sentence in which they appear. When adjustments are in tables, the (R) follows the
designated adjustment. Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
2