TB 9-6625-2097-35
NOTE
Unless otherwise specified, verify the result of each test and,
whenever the test requirement is not met, take corrective
action before continuing with the calibration. Adjustments
required to calibrate the TI are included in this procedure.
Additional maintenance information is contained in the
manufacturer's manual for this TI.
8. Channel Span Gain
a. Performance Check
(1) Connect dc voltage standard (A3) to SIGNAL INPUT I (fig. 2), using cable (B2).
Figure 2. Model 2200S (Series) - rear view
6