TB 9-6625-2117-24
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Oscilloscope, Tektronix, Type 434. The manufacturer's manual was used as
the prime data source in compiling these instructions. The equipment being calibrated will
be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. Adjustments vary with serial numbers which are indicated in text.
b. Time and Technique. The time required for this calibration is approximately 3
hours, using the dc and low frequency technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables the (R) follows the designated adjustments.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
Power input requirement
115 V ac at 60 Hz
Calibrator output
Range: 0.6 V at 1000 Hz
1%
Accuracy:
Range: 1 mV to 10 V/div
3%
Accuracy:
Risetime
14 ns or less
Range: 5 s/div to 0.2 s/div
Sweeptime
Unmag 3%, mag 4%
Accuracy:
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286, AN/GSM-287 and AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI. Where the four-to-one ratio
cannot be met, the actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The accessories required for this calibration are common
usage accessories issued as indicated in paragraph 4 above and are not listed in this