TB 9-6625-2124-24
(9) Press function keys and enter corresponding data on keyboard as listed in (a)
through (e) below:
INSTR PRESET key.
(a)
START key, DATA ENTRY 10 MHz.
(b)
STOP key, DATA ENTRY 2 GHz.
(c)
M1 key, DATA ENTRY 1 GHz.
(d)
M2 key, DATA ENTRY 1.8 GHz.
(e)
(10) Press AMPTD MKR (sweep oscillator) key (ON) and ensure plug-in AMPTD
MKR is off.
(11) Repeat (3) above.
(12) Adjust oscilloscope position control to display M1 (first) marker on crt,
increasing oscilloscope horizontal timing setting as necessary.
leading of M1 marker on oscilloscope display. Synthesized signal generator frequency will
indicate between 0.975 and 1.025 GHz.
(14) Repeat technique of (12) and (13) above for M2 marker. Synthesized signal
generator frequency will indicate between 1.775 and 1.825 GHz.
(15) Press function keys and enter corresponding data on keyboard as listed in (a)
through (d) below:
START key, DATA ENTRY 2 GHz.
(a)
STOP key, DATA ENTRY 8.4 GHz.
(b)
M1 key, DATA ENTRY 2.5 GHz.
(c)
M2 key, DATA ENTRY 8 GHz.
(d)
(16) Repeat (3) above.
(17) Repeat (12) above.
(18) Repeat technique of (13) above. Synthesized signal generator frequency will
indicate between 2.448 and 2.552 GHz.
(19) Repeat technique of (12) and (13) above for M2 marker. Synthesized signal
generator frequency will indicate between 7.948 and 8.052 GHz.
(20) Repeat (3) and (5) above.
(22) Repeat (7) above.
b. Adjustments
for multimeter indication of -10.000 0.001 V dc (R).
(2) Connect multimeter to A6TP13 (+) and A6TP16 (-) (fig. 3).
(3) Press function keys and enter corresponding data on keyboard as listed in (a)
through (g) below:
14