Figure 4. Swept frequency and marker accuracy - equipment setup.
Swept Frequency and Marker Accuracy - Continued

Calibration Procedure For Sweep Oscillator, Modles 8350A, 8350B, And Plug-In Models 8325A, 83592A, And 83592A Opt 002 Page Navigation
  9    10    11    12    13  14  15    16    17    18    19  
img
. . .
TB 9-6625-2124-24
(9) Press function keys and enter corresponding data on keyboard as listed in (a)
through (e) below:
INSTR PRESET key.
(a)
START key, DATA ENTRY 10 MHz.
(b)
STOP key, DATA ENTRY 2 GHz.
(c)
M1 key, DATA ENTRY 1 GHz.
(d)
M2 key, DATA ENTRY 1.8 GHz.
(e)
(10) Press AMPTD MKR (sweep oscillator) key (ON) and ensure plug-in AMPTD
MKR is off.
(11) Repeat (3) above.
(12) Adjust oscilloscope position control to display M1 (first) marker on crt,
increasing oscilloscope horizontal timing setting as necessary.
(13) Adjust synthesized signal generator frequency to align the birdie with the
leading of M1 marker on oscilloscope display. Synthesized signal generator frequency will
indicate between 0.975 and 1.025 GHz.
(14) Repeat technique of (12) and (13) above for M2 marker. Synthesized signal
generator frequency will indicate between 1.775 and 1.825 GHz.
(15) Press function keys and enter corresponding data on keyboard as listed in (a)
through (d) below:
START key, DATA ENTRY 2 GHz.
(a)
STOP key, DATA ENTRY 8.4 GHz.
(b)
M1 key, DATA ENTRY 2.5 GHz.
(c)
M2 key, DATA ENTRY 8 GHz.
(d)
(16) Repeat (3) above.
(17) Repeat (12) above.
(18) Repeat technique of (13) above. Synthesized signal generator frequency will
indicate between 2.448 and 2.552 GHz.
(19) Repeat technique of (12) and (13) above for M2 marker. Synthesized signal
generator frequency will indicate between 7.948 and 8.052 GHz.
(20) Repeat (3) and (5) above.
(21) Synthesized signal generator frequency will indicate between 1.980 and 2.020 GHz.
(22) Repeat (7) above.
(23) Synthesized signal generator frequency will indicate between 8.380 and 8.420 GHz.
b. Adjustments
(1) Connect multimeter to A6TP3 (+) and A6TP5 (-) (fig. 3). Adjust A6R21 (fig. 3)
for multimeter indication of -10.000 0.001 V dc (R).
(2) Connect multimeter to A6TP13 (+) and A6TP16 (-) (fig. 3).
(3) Press function keys and enter corresponding data on keyboard as listed in (a)
through (g) below: