Figure 2. CW frequency accuracy equipment setup.
Figure 4. Swept frequency and marker accuracy - equipment setup.

Calibration Procedure For Sweep Oscillator, Modles 8350A, 8350B, And Plug-In Models 8325A, 83592A, And 83592A Opt 002 Page Navigation
  7    8    9    10    11  12  13    14    15    16    17  
img
. . . . . .
TB 9-6625-2124-24
b. Adjustments
(1) Press INSTR PRESET and STOP keys. Enter data 2 GHz on keyboard.
(2) Press TI (plug in) AMPTD MKR key (ON). Connect microwave frequency
counter to A8TP1 (fig. 3).
(3) Adjust A8C4 (fig. 3) for microwave frequency counter indication of 50,000,000
250 Hz (R).
Figure 3. Model 83525A plug-in - adjustment locations.
14. Swept Frequency and Marker Accuracy
a. Performance Check
(1) Connect equipment as shown in figure 4.
(2) Press function keys and enter corresponding data on keyboard as listed in (a)
through (d) below:
(a) INSTR PRESET key.
12