TB 9-6625-2124-24
Figure 2. CW frequency accuracy equipment setup.
(2) Press function keys and enter corresponding data on keyboard as listed in (a)
through (c) below:
(a) INSTR PRESET key.
(b) CW key, DATA ENTRY 50 MHz.
(c) POWER LEVEL key, DATA ENTRY 10 dBm.
(3) Adjust FREQ CAL for microwave frequency counter indication of 50.00 MHz.
between 5.0 and 15.0 MHz; if not, perform b below.
indicate as specified; if not, perform b below.
Test
Frequency counter
instrument
indications (GHz)
data entry
Min
Max
1
GHz
0.995
1.005
2
GHz
1.995
2.005
2.1
GHz
2.085
2.115
6.0
GHz
5.985
6.015
8.4
GHz
8.385
8.415
indicate within 20 kHz of value recorded in (6) above.
indicate within 20 kHz of value recorded in (6) above.
(9) Adjust autotransformer for a 115 V ac output.
11