TB 9-6625-2124-24
Figure 1. Test instrument mainframe - adjustment locations.
(10) Press function key and enter corresponding data on keyboard as listed in (a)
through (e) below:
INSTR PRESET key.
(a)
CW key, DATA ENTRY -2000 MHz.
(b)
SHIFT key DATA. ENTRY 00.
(c)
M1 key, DATA ENTRY 3000.
(d)
M2 key, DATA ENTRY 01.
(e)
(11) The FREQUENCY/TIME display will indicate 01. Adjust A4R44 (fig. 1) for
FF. Adjust A4R22 (fig. 1) for multimeter indication of -0.19492 V dc (R).
(14) Press function keys and enter corresponding data on keyboard as listed in (a)
and (b) below:
(a) INSTR PRESET key.
(b) CW key, DATA ENTRY 10.2 GHz.
(15) Adjust A4R27 (fig. 1) for multimeter indication of 10.2000 V dc (R).
(16) Press function keys and enter corresponding data on keyboard as listed in (a)
through (c) below: