Table 1. Calibration Description
Table 2. Minimum Specifications of Equipment Required

Calibration Procedure For Sweep Oscillator, Modles 8350A, 8350B, And Plug-In Models 8325A, 83592A, And 83592A Opt 002 Page Navigation
  1    2    3  4  5    6    7    8    9  
img
TB 9-6625-2124-24
Table 1. Calibration Description - Continued
Models 83592A and 83592A Opt 002
0.01 to 20 GHz1
Markers
Range:
Accuracy:
15 MHz ( 0.5% sweep width)
0.01 to 2.4 GHz
20 MHz ( 0.5% sweep width)
2.4 to 7 GHz
25 MHz, ( 0.5% sweep width)
7 to 13.5 GHz
30 MHz ( 0.5% sweep width)
13.5 to 20 GHz1
Leveled
Range:
-5 to +10 dBm
output power
Accuracy:
1.5 dB, 0.01 to 2.4 GHz
(83592A)
1.3 dB, 2.4 to 13.5 GHz
1.4 dB, 13.5 to 18.6 GHz1
Leveled
Range:
-5 to +10 dBm, 0.01 to 2.4 GHz
output power
-5 to 8.5 dBm, 2.4 to 7 GHz
(83592A
-5 to +8 dBm, 7 to 13.5 GHz
-5 to +7 dBm, 13.5 to 18.6 GHz1
opt 002)
Accuracy:
1.7 dB, 0.01 to 2.4 GHz
1.5 dB, 2.4 to 13.5 GHz
1.6 dB, 13.5 to 18.6 GHz1
0.01 to 20 GHz1
Internal
Range:
power
Accuracy:
0.9 dB, 0.01 to 2.4 GHz
leveling
0.7 dB, 2.4 to 13.5 GHz
0.8 dB, 13.5 to 20 GHz1
0.01 to 20 GHz1
Power meter
Range:
leveling
Accuracy:
0.2 dB
Step
Range:
-10 to -70 dB
attenuator
(83592A
Opt 002)
Accuracy:
Attenuator setting (dB)
10
20
30
40
50
60
70
Frequency range
0.01 to
12.4 GHz
2.0
tolerance in dB
0.6
0.7
0.9
1.8
2.2
2.3
Frequency range
12.4 to
18 GHz
tolerance in dB
0.7
0.9
1.2
2.0
2.3
2.5
2.8
Models 83592A and 83592A Opt 002
AM ON/OFF ration
Range:
>30 dB at 1 and 4 GHz with +10 dBm output level
Square wave
Range:
40/60% (ON/OFF) at +10 dBm
symmetry
1Calibrated