TB 9-6625-2124-24
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument
parameters
Performance specifications
Models 8350A and 8350B
Tuning voltage
Range:
0, 5, and 10 V dc
Accuracy:
0.04%
Model 83525A
Range:
0.01 to 8.4 GHz
Accuracy:
5 MHz, 0.01 to 2 GHz
15 MHz, 2 to 8.4 GHz
20 kHz at 10% line charge
Frequency
Accuracy:
stability
Swept
Range:
0.01 to 8.4 GHz
Accuracy:
15 MHz, 0.01 to 2 GHz
2 20 MHz, 2 to 8.4 GHz
Markers
Range:
0.01 to 8.4 GHz
Accuracy:
15 MHz ( 0.5% sweep width)
0.01 to 2 GHz
20 MHz ( 0.5% sweep width)
2 to 8.4 GHz
Leveled
Range:
-2 to +13 dBm
output power
Accuracy:
1.5 dBm
Power
Range:
0.01 to 8.4 GHz
leveling
Accuracy:
AM ON/OFF ratio
Range:
>30 dB at 1 and 4 GHz with +13 dBm output level
Square wave
Range:
40/60% (ON/OFF) at +13 dBm
symmetry
Models 83592A and 83592A Opt 002
0.01 to 20 GHz1
Range:
Accuracy:
5 MHz, 0.01 to 7 GHz
10 MHz, 7 to 13.5 GHz
15 MHz, 13.5 to 20 GHz1
Frequency
Accuracy:
50 kHz at 10% line change
stability
0.01 to 20 GHz1
Swept
Range:
50 MHz, 0.01 to 20 GHz1
Accuracy:
15 MHz, 0.01 to 2.4 GHz
20 MHz, 2.4 to 7 GHz
25 MHz, 7 to 13.5 GHz
30 MHz, 13.5 to 20 GHz1
See footnote at end of table.