• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Table 1. Calibration Description - Continued
Equipment Setup

Calibration Procedure For Sweep Oscillator, Modles 8350A, 8350B, And Plug-In Models 8325A, 83592A, And 83592A Opt 002 Page Navigation
  1    2    3    4  5  6    7    8    9    10  
img
. . . . . . . . .
TB 9-6625-2124-24
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure.  This equipment is issued with Secondary Transfer Calibration
Standards Sets AN/GSM-287, or AN/GSM-705.  Alternate items may be used by the
calibrating activity. The items selected must be verified to perform satisfactorily prior to
use and must bear evidence of current calibration. The equipment must meet or exceed the
minimum use specifications listed in table 2. The accuracies listed in table 2 provide a four-
to-one ratio between the standard and TI. When the four-to-one ratio cannot be met, the
actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required.  The accessories required for this calibration are common
usage accessories issued as indicated in paragraph 4 above and are not listed in this
calibration procedure.  The following peculiar accessories are also required for this
calibration:  Directional Coupler, Narda, Model 3095 (3095); and Semiconductor Device
(Crystal Detector), Hewlett-Packard, Model 423AOPT03 (7923182).
Table 2. Minimum Specifications of Equipment Required
Common name
Minimum use
Manufacturer and model
(official nomenclature)
specifications
(part number)
ATTENUATOR (FIXED)
Range:
10 dB
Weinschel, Model 9918-10dB
(9918-10dB)
Accuracy: Determined in
procedure
AUTOTRANSFORMER
Range: 108 to 125 V ac
Ridge, Model 9020A (9020A)
1%
Accuracy:
FREQUENCY COUNTER
Range:
10.0 ms to 50 MHz
Fluke, Model PM6681/656
0.05%
(PM6681/656)
Accuracy:
MEASURING RECEIVER
Range: 0 to -70 dB
Measuring receiver system
7.5%
N5530S consisting of: Spectrum
Accuracy:
Analyzer Agilent, Model E4440A
(E4440A), Power meter Agilent,
Range: 0.850 to 18.0 GHz
0.0625%
model E4419B (E4419B), and
Accuracy:
Sensor module Agilent, Model
N5532A opt 518 (518)
MICROWAVE FREQUENCY
Range:
5 MHz to 18 GHz
Anritsu, Model MF2414B003
5 X 10-6
COUNTER
(MF2414B003)
Accuracy:
MULTIMETER
Range:
-40 to +20.0 V dc
Hewlett-Packard, Model 3458A
0.01%
(3458A)
Accuracy:
OSCILLOSCOPE
Range:
Dc to 10 ms
Agilent, OS-303/G (OS-303/G)
POWER METER
Range: -6.5 to + 10 dBm
Hewlett-Packard, Model 437B
Accuracy:  6%
(13440045)
w/power sensor,
Hewlett-Packard, Model
8485A
SPECTRUM ANALYZER
Range:
1 to 4 GHz
(AN/USM-677)
Anritsu, Model 68369NV
SYNTHESIZED SIGNAL
Range:
2 to 8 GHz
(68369NV)
GENERATOR
Accuracy:
3.75 MHz (
0.125%)
5


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.