Figure 1. Test instrument mainframe - adjustment locations.
Equipment Setup

Calibration Procedure For Sweep Oscillator, Modles 8350A, 8350B, And Plug-In Models 8325A, 83592A, And 83592A Opt 002 Page Navigation
  4    5    6    7    8  9  10    11    12    13    14  
img
. . . . . . . . . . . . .
TB 9-6625-2124-24
(a) INSTR PRESET key.
(b)  F key, DATA ENTRY 10.4 GHz.
(c) SWEEP MAN key, DATA ENTRY 10.2 GHz.
(17) Adjust A4R25 (fig. 1) for multimeter indication of 10.2000 V dc (R).
(18) Enter data -200 MHz on keyboard. Adjust A4R2 (fig. 1) for multimeter
indication of -0.2000 V dc (R).
(19) Press F key and enter data 0 GHz on keyboard. If multimeter does not
indicate 5.00000 V dc, press VERNIER key and adjust FREQUENCY VERNIER control
for 5.00000 V dc multimeter indication.
(20) Press F key and enter data 1.299 GHz. Adjust A4R15 (fig. 1) for multimeter
indication of 4.35064 V dc (R).
(21) Press F key and enter data 162.4 MHz on keyboard. Adjust A4Rl8 (fig. 1) for
multimeter indication of 4.91883 V dc (R).
9. Power Supply
NOTE
Do not perform power supply checks if all other parameters are
within tolerance.
NOTE
Install plug-in in TI prior to performing this check.
a. Performance Check
(1) Connect multimeter to A7TP16 (-) and A7TP15 (+) (fig. 1). If multimeter does
not indicate between 19.990 and 20.010 V dc, perform b (1) below.
(2) Disconnect multimeter from A7TP16 and A7TP15. Connect to A7TP17 (-) and
A7TP18 (+) (fig. 1). If multimeter does not indicate between -39.980 and -40.020 V dc,
perform b (2) below.
(3) Disconnect multimeter from A7TP17 and A7TP18. Connect to A7TP8 (-) and
A7TP9 (+) (fig. 1). If multimeter does not indicate between -9.995 and -10.005 V dc,
perform b (3) below.
b. Adjustments
(1) Adjust A7R10 (fig. 1) for multimeter indication of 20.000 V dc (R).
(2) Adjust A7R39 (fig. 1) for multimeter indication of -40.000 V dc (R).
(3) Adjust (fig. 1) for multimeter indication of -10.000 V dc (R).
10. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
9