TB 9-6625-2142-40
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
Range: <-10 to > + 10 V dc
Hewlett-Packard, Model 3458A
Accuracy: .05%
(3458A)
OSCILLOSCOPE
Voltage output:
Fluke, Model 5820A-5C-GHZ
(5820A-5C-GHZ)
CALIBRATOR
Range: 0.1 to 2 V at 1 and 10 kHz
Accuracy: 0.75%
Time markers:
Range: 5 ms to 5 ns
Accuracy: 0.75%
OSCILLOSCOPE NO. 1
Compatible with sampling sweep unit
Tektronix, Type 7613 (MIS-
and sampling unit
35808/1)
Range: 1 ms/div to 0.05 s/div
(OS-303/G)
OSCILLOSCOPE NO. 2
1 V/div to 0.2 mV/div
Accuracy: 1.4% ( 4%)
SAMPLING HEAD
Compatible with sampling unit
Tektronix, Type S-4 (MIS-35808/2)
SAMPLING SWEEP
Tektronix, Types 7T11 or 7T11A
UNIT
sampling unit
(MIS-35808/6)
SAMPLING UNIT
Tektronix, Type 7S11
and sampling sweep unit
(MIS-35808/4)
(TEKTRONIX, TYPE 7S11)
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.