Horizontal Timing
Figure 5. Outside board - right side.

Calibration Procedure For Sampling Unit Am-6787/U (Type 7S11) And Sampling Sweep Unit Td-1161/U (Types 7T11 And 7T11A) Page Navigation
  8    9    10    11    12  13  14    15  
img
. . . . . .
TB 9-6625-2142-40
Table 5. Sweep Timing
Test instrument
Oscilloscope
Oscilloscope
Adjustments
SWEEP RANGE
calibrator
markers
(fig. 4)
switch settings
settings
per division
(R)
5  ms
0.5 ms
1
---
s
0.5 ms
1
R260
50
50  ns1
10  ns
2
C292
0.5  s
50  ns
1
---
s
0.5  s
1
---
5
s
s
1
---
50
5
1Press
TI HF SYNC pushbutton. Adjust TI STABILITY control for stable display. Check at both RANDOM and
SEQUENTIAL pushbutton settings.
(14) Set TIME/DIV switch to .5 s and press TI SEQUENTIAL pushbutton.
(15) Set oscilloscope calibrator for 1 s. Crt will display 1 marker for each 2 divisions.
Figure 4. Interior board - right side.
(16) Release VARIABLE (CAL IN) control to out position and turn fully cw. Crt
will display markers at least 5 divisions apart.
(17) Press VARIABLE (CAL IN) control to in position.
b. Adjustments. Adjust oscilloscope calibrator for a 5 ms output. Adjust R265 (fig. 4)
for 1 marker per division over center 8 divisions (R).
14. Triggering