Figure 5. Outside board - right side.

Calibration Procedure For Sampling Unit Am-6787/U (Type 7S11) And Sampling Sweep Unit Td-1161/U (Types 7T11 And 7T11A) Page Navigation
  10    11    12    13    14  15
img
. . . . . .
TB 9-6625-2142-40
(2) Turn TRIG LEVEL and STABILITY controls fully ccw and adjust R10 (fig. 5)
for a 4.5 ns period (R).
15. Power Supply
a. Performance Check. Set digital multimeter to read 20 V dc. Connect multimeter
to TP602 (fig. 4) and ground. If multimeter does not indicate between +9.908 and +10.02 V
dc, perform b below.
b. Adjustments. Adjust R600 (fig. 4) for an indication of 10.00 V dc (R).
16. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
15/(16 Blank)