TB 9-6625-2155-24
Table 2. Minimum Specifications of Equipment Required - Continued
Minimum use
Manufacturer and model
Common name
specifications
(part number)
MEASURING RECEIVER
Range: 2.0 to 18 GHz
Measuring receiver system
Range: +8 to -74.5 dBm
N5530S consisting of:
Accuracy: +.5 dB
Spectrum Analyzer, Agilent
Range: AM 0 to 80%
Model E4440A (E4440A),
Accuracy: 2% at 1 kHz
Power meter, Agilent Model
E4419B (E4419B), and
Range: FM .05 to 100 kHz
Accuracy: 2% at 1 kHz
Sensor module, Agilent Model
Deviation: 12 kHz
N5532A opt. 504 (504), 518
(518), 526 (526)
Range: 0 to 40 V dc
Fluke, Model 8840A/AF 05
Accuracy: 0.01%
(AN/GSM-64D)
Range: 5.0 V at 100 ns
(OS-303/G)
Accuracy: 3.0%
Range: Period 10 s
LeCroy, Model 9210MOD200
(9210MOD200) w/plug-ins,
LeCroy, Models 9211 (9211)
and 9215 (9215) (MIS 45839)
SYNTHESIZED SIGNAL
Range: Output level +8 dBm
Anritsu, Model 68369NV
(AN/USM-677)
SPECTRUM ANALYZER
Range: 2.0 to 18 GHz at +10 to 60 to dBm
of the span/div
Range: Span 500 Hz to 20 MHz
Accuracy: 5%
Datum, Model ET6000-75
TIME/FREQUENCY
Frequency: 1 MHz
-10 per day
(13589305)
WORKSTATION
Accuracy: 5 parts in 10
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual and TM 11-6625-3143-
40 for this TI.
d. When indications specified in paragraphs 8 through 14 are not within tolerance,
perform the power supply check prior to making adjustments. After adjustments are made,