TB 9-6625-2155-24
(9) Press TI keys as listed in (a) through (f) below.
RCL 0.
(a)
Adjust OUTPUT LEVEL for 0 dB.
(b)
ALC INTERNAL on.
(c)
RF OUTPUT on.
(d)
FREQ INCR, 1, 1, 1, ., 1, 1, 3, and MHz.
(e)
(f)
(10) With the measuring receiver still configured for a frequency measurement,
verify that the measuring receiver indicates within tolerances listed in table 4.
(11) Repeat (9) (e) and (10) for remaining settings in table 4 verifying measuring
receiver indicates within the tolerances listed.
Test instrument
Measuring receiver indication
Step
Step
Frequency
increment
increment
(GHz)
setting
direction
Minimum
Maximum
2
111.113
1999.99700M
2000.00300M
UP
2111.11000M
2111.11600M
UP
2222.22300M
2222.22900M
UP
2333.33600M
2333.34200M
UP
2444.44900M
2444.45500M
UP
2555.56300M
2555.56800M
UP
2666.67600M
2666.68100M
UP
2777.78800M
2777.94000M
UP
2888.90100M
2888.90700M
2
1000
UP
2999.99700M
3000.00300M
UP
3999.99700M
4000.00300M
UP
4999.99700M
5000.00300M
UP
5999.99700M
6000.00300M
UP
6999.99700M
7000.00300M
UP
7999.99700M
8000.00300M
UP
8999.99700M
9000.00300M
UP
9999.99700M
10000.00300M
UP
10999.99700M
11000.00300M
UP
11999.99700M
12000.00300M
UP
12999.99700M
13000.00300M
UP
13999.99700M
14000.00300M
UP
14999.99700M
15000.00300M
UP
15999.99700M
16000.00300M
UP
16999.99700M
17000.00300M
UP
17999.99700M
18000.00300M
(12) Reduce all outputs to minimum and disconnect equipment.
b.
Adjustments. No adjustments can be made.
7