Table 7. Non-harmonic Spurious Signal Level
Table 12. 12.5 GHz Output Level Test

Calibration Procedure For Signal Generator Sg-1219/U (Model 8673M) Page Navigation
  6    7    8    9    10  11  12    13    14    15    16  
img
. . . .
TB 9-6625-2155-24
(4) Using measuring receiver, configured for an RF power measurement, sweep the
TI from 2 GHz to 18 GHz in 1 GHz steps, and record the highest and lowest levels.
(5) Calculate the flatness using the formula below.  If the flatness is not less than
or equal to the maximum limit listed in table 8, perform b below
Flatness = (highest lowest) / 2
Table 8. Output Level Flatness
Start frequency
Stop frequency
Max limit
2 GHz
18 GHz
1.999
(6) Press TI FREQUENCY, 2, and GHz keys. Using tuned level measurement
techniques, verify that the measuring receiver indicates within minimum and maximum
limits for TI output levels listed in table 9.
Table 9. 2 GHz Output Level Test
Test
instrument
output level
Min
Max
0
-2
2
-10
-12
-8
-20
-22
-18
-30
-32
-28
-40
-42
-38
-50
-52
-48
-60
-62
-58
-70
-73.5
-66.5
-80
-83.5
-76.5
-90
-93.5
-86.5
-100
-103.5
-96.5
(7) Press TI FREQUENCY, 6, and GHz keys. Using tuned level measurement
techniques, verify that the measuring receiver indicates within minimum and maximum
limits for TI output levels listed in table 10.
Table 10. 6 GHz Output Level Test
Test
instrument
output level
Min
Max
0
-2
2
-10
-12
-8
-20
-22
-18
-30
-32
-28
-40
-42
-38
-50
-52
-48
-60
-62
-58
-70
-73.5
-66.5
-80
-83.5
-76.5
-90
-93.5
-86.5
-100
-103.5
-96.5
12