TB 9-6625-2155-24
(f) AUTO PEAK on.
(g) PULSE NORMAL on.
(4) Press synthesized signal generator as listed in (a) through (e) below.
RCL 0.
(a)
Adjust OUTPUT LEVEL for 8 dB.
(b)
ALC INTERNAL on.
(c)
RF OUTPUT on.
(d)
(e)
(5) Set the pulse generator to produced the outputs listed in (a) through (f) below.
Recall Setup, Standard, Enter, Enter.
(a)
(b)
Frequency to 1 MHz.
A:Width to 110 nS.
(c)
A:VHI to 5 V.
(d)
A:VLO to 0 V.
(e)
A:DISP off.
(f)
(6) Set the oscilloscope as listed in (a) through (i) below.
Vertical, 1 and 2 on (lit).
(a)
Vertical, 1 and 2 Input to 50 (lit).
(b)
Vertical 1, scale to 20 mV.
(c)
Vertical 2, scale to 100 mV.
(d)
Trigger Sweep to Auto.
(e)
Trigger Source to 2.
(f)
Trigger Coupling to DC.
(g)
.
Trigger Slope to
(h)
Horizontal sweep speed to 200 ns.
(i)
NOTE
The next several tests require that you adjust the oscilloscope
as necessary to obtain a centered 5 division pulse as shown in
17