• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Pulse Modulation
Table 14. Pulse Modulation Overshoot and Undershoot Check - Continued

Calibration Procedure For Signal Generator Sg-1219/U (Model 8673M) Page Navigation
  12    13    14    15    16  17  18    19    20    21    22  
img
. . .
TB 9-6625-2155-24
Figure 8. Pulse modulated display.
(7) Verify pulse overshoot and undershoot is less than 20 percent and ALC
UNLEVEL light on TI is extinguished. If signal is not within limits specified perform
adjustments listed in b below.
(8) Set TI, and synthesized signal generator to the remaining frequencies and levels
listed in table 14, and repeat (7) above at each frequency.
Table 14. Pulse Modulation Overshoot and Undershoot Check
Test instrument
Synthesized
Frequency
Level
signal
(GHz)
(dB)
generator
Description
frequency
(GHz)
2.00
3
2.050
Overshoot Undershoot @ 2.0 GHz
+3dBm
2.00
-10
2.050
Overshoot Undershoot @ 2.0 GHz
-10dBm
6.60
3
6.650
Overshoot Undershoot @ 6.6 GHz
+3dBm
6.60
-10
6.650
Overshoot Undershoot @ 6.6 GHz
-10dBm
6.60
0
6.650
Overshoot Undershoot @ 6.6 GHz
0dBm
6.70
3
6.750
Overshoot Undershoot @ 6.7 GHz
+3dBm
6.70
-10
6.750
Overshoot Undershoot @ 6.7 GHz
-10dBm
6.70
0
6.750
Overshoot Undershoot @ 6.7 GHz
0dBm
12.290
3
12.340
Overshoot Undershoot @ 12.29 GHz
+3dBm
12.290
-10
12.340
Overshoot Undershoot @ 12.29 GHz
-10dBm