TB 9-6625-2155-24
9. Single Sideband Phase Noise
a. Performance Check
(2) Press TI keys as listed in (a) through (e) below.
RCL 0.
(a)
Adjust OUTPUT LEVEL for 8 dB.
(b)
ALC INTERNAL on.
(c)
RF OUTPUT on.
(d)
(e)
(3) Set spectrum analyzer as listed in (a) through (f).
Preset.
(a)
AMPLITUDE, Ref Level to 8 dBm.
(b)
(c)
All markers off.
(d)
BW, Video BW, and Res BW to Auto.
(e)
SPAN to 1 MHz.
(f)
(4) Allow spectrum analyzer to sweep signal a couple of times then set spectrum
analyzer as listed in (a) through (c) below:
(a) FREQUENCY, Signal Track to On.
(b) SPAN to 2.5 kHz.
(c) BW/Avg, Video BW to 10 Hz.