TB 9-6625-2162-24
(6) Set DISPLAY SELECT switch to PW (MICROSEC) and press 5 and then E on
keyboard.
(7) Set DISPLAY SELECT switch to RF SIG PWR (DBM) and adjust
ATTENUATION DB FINE control for a display indication of 0.0.
(8) Adjust oscilloscope vertical position control until peaks of displayed pulses
are displayed.
NOTE
Pulse baseline will be off the crt display.
(9) Set MODE switch to SIG GEN CW and adjust ATTENUATION DB FINE
control to position oscilloscope trace on graticule centerline.
(10) Disconnect semiconductor device from TI and connect measuring receiver to RF
IN/OUT connector and configure measuring receiver for a power measurement. Measuring
receiver will indicate 0 1 dBm.
(11) Disconnect measuring receiver from TI and reconnect semiconductor device to
RF IN/OUT connector.
(12) Set DISPLAY SELECT switch to PW (MICROSEC), MODE switch to SIG
GEN PULSE, and press 1 and then E on keyboard. Repeat technique of (7) through
(11) above.
(13) Set DISPLAY SELECT switch to PW (MICROSEC), MODE switch to SIG
GEN PULSE, and press 0.2 and then E on keyboard. Repeat technique of (7) through
(11) above.
b. Adjustments. No adjustments can be made.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.