Figure 7. Pulse characteristics connection.
PRF and Pulse Characteristics

Calibration Procedure For Radar Test Set, Model An/Upm-145 Page Navigation
  14    15    16    17    18  19  20    21    22    23  
img
. . . .
TB 9-6625-2162-24
12. Target Rate and Range
a. Performance Check
(1) Connect equipment as shown in figure 8.
Figure 8. Target rate and range connection.
(2) Position controls as listed in (a) through (e) below:
DISPLAY SELECT switch to PRF (Hz).
(a)
MODE switch to SIG GEN PULSE.
(b)
TRIGGER switch to IN POS.
(c)
ATTENUATION DB 0-80 switch to 0.
(d)
ATTENUATION DB FINE control fully cw.
(e)
(3) Set function generator controls for a 1Ps pulse at an amplitude of 1 V and a
period of 5 sec.
(4) Adjust pulse generator controls for a positive 1 Ps pulse at an amplitude of 10 V.
Adjust pulse generator frequency controls for a TI display indication of 1000.
(5) Set DISPLAY SELECT switch to PW (MICROSEC) and press 1.0 and then E
on keyboard.