TB 9-6625-2162-24
(6) Position controls as listed in (a) through (d) below:
MODE switch to SIG GEN PULSE.
(a)
(b)
(c)
DISPLAY SELECT switch to PW (MICROSEC).
(d)
(7) Press 5 and then E on keyboard. Set oscilloscope controls to measure pulse
width. Oscilloscope will indicate within limits specified in first row of table 7.
(8) Repeat (7) above using keyboard entries listed in table 7.
Oscilloscope will
indicate within limits specified in table 7.
Test instrument
keyboard entries
Min
Max
5 and E
4.85
5.15
4 and E
3.88
4.12
3 and E
2.91
3.09
2 and E
1.94
2.06
1 and E
0.95
1.05
0.5 and E
0.45
0.55
0.2 and E
0.15
0.25
(9) Press 5 and then E on keyboard. Measure the pulse rise time. Rise time will be
0.1 Ps or less.
(10) Measure pulse decay time. Decay time will be 0.1 Ps or less.
(11) Repeat (9) and (10) above for keyboard entries of 4, 3, 2, and 1.
b. Adjustments. No adjustments can be made.
19