Figure 6. Main board.
Figure 8. Oscilloscope  left side view.

Calibration Procedure For Oscilloscope Types Sc501, Sc502, And Sc504 Page Navigation
  15    16    17    18    19  20  21    22    23    24    25  
img
. . . .
TB 9-6625-2173-24
R3110
R3115
R3466
TP3465
R3336
Figure 7. Trigger board.
28. Vertical Gain
a. Performance Check
(1) Set TI SECONDS/DIV switch to .5 ms.
(2) Set TI CH 1 and CH2 VOLTS/DIV switches to 10m.
(3) Connect oscilloscope calibrator SOURCE/MEASURE CHAN 1 to TI CH 1 and
CHAN 2 to TI CH 2.
(4) Set oscilloscope calibrator VOLTAGE output to 50 mV, 1 kHz.
(5) Adjust TRIGGERING LEVEL and POSITION controls, as necessary, to
view waveform.
(6) Adjust INTEN and FOCUS controls for suitable viewing.
(7) Rotate oscilloscope calibrator EDIT FIELD knob for 5 divisions of vertical
deflection.  If oscilloscope calibrator err display readout does not indicate within r2%,
perform b (1).
(8) Repeat technique of (1) through (7) above for settings listed in table 5.
Oscilloscope calibrator err display readout will indicate within r2%.
(9) Set TI DISPLAY and SOURCE switches to CH 2.
(10) Activate oscilloscope calibrator CHAN 2 output and repeat technique of (1)
through (7) above for CH 2. If oscilloscope calibrator err display readout does not indicate
within limits specified, perform b (2).