Equipment Setup
Figure 7. Trigger board.

Calibration Procedure For Oscilloscope Types Sc501, Sc502, And Sc504 Page Navigation
  14    15    16    17    18  19  20    21    22    23    24  
img
. . . .
TB 9-6625-2173-24
(3) Adjust  TRIGGERING  LEVEL,  CH1
HORIZONTAL
POSITION
and
TIME/DIV controls for a stable, centered display.
(4) Adjust CH1 VOLTS/DIV and VAR control for 6 divisions of vertical
deflection on TI.
(5) Remove connection at TI CALIBRATOR and connect oscilloscope calibrator
SOURCE/MEASURE CHAN 1 to TI CH1 (do not change setting set in (4) above).
(6) Set oscilloscope calibrator VOLTAGE output to 600 mV, 1 kHz.
(7) Rotate oscilloscope calibrator EDIT FIELD knob to adjust for 6 divisions of
vertical deflection on TI. Oscilloscope calibrator err display will indicate within r1%, if not,
perform b below.
b. Adjustments
(1) Connect multimeter HI to TI CAL 0.6V and LO to chassis ground.
(2) Short TP1110 and TP1112 (fig. 6).
(3) Adjust R3466 (fig. 7) for a 0.6000 V dc indication on multimeter. (R).
(4) Disconnect short from between TP1110 and TP1112 (fig. 6).
(5) Disconnect multimeter from TI.
Figure 6. Main board.
19