Phase Modulation
Power Supply

Calibration Procedure For Signal Generator Sg-1207/U (Model 8642M) Page Navigation
  11    12    13    14    15  16  17    18  
img
. . . . . . .
TB 9-6625-2182-24
Table 20. PM Accuracy
Measuring receiver phase
Test instrument
Test instrument
DATA carrier
DATA phase
modulation indications (rad)
frequency
modulation
Min
Max
8 MHz
75 rad
71.2
78.8
1050 MHz
100 rad
94.9
105.1
500 kHz
100 rad
94.9
105.1
(4) Perform steps (a) through (c) below for each row in table 21.
(a) Press ENTRY FREQ and enter DATA carrier frequency.
(b) Press SHIFT -  M and enter DATA rad.
(c) Using measuring receiver, measure the audio distortion.
Distortion will
indicate within limits specified.
Table 21. Audio Distortion at 1 kHz Modulation Frequency
Test instrument
Test instrument
Measuring receiver
DATA
DATA
distortion indications
carrier frequency
phase modulation
(< %)
8 MHz
75  rad
0.4
1050 MHz
100  rad
0.4
500 kHz
100  rad
0.4
(5) Press RF OFF/ON pushbutton to OFF.
(6) Disconnect measuring receiver from TI OUTPUT RF.
b. Adjustments. No adjustments can be made.
18. Internal Oscillator
a. Performance Check
(1) Connect TI OUTPUT MOD to audio analyzer INPUT HIGH.
(2) Press TI pushbuttons as listed in (a) through (d) below:
INSTR PRESET.
(a)
SHIFT.
(b)
ENTRY - MOD OUT.
(c)
DATA - (+1 V).
(d)
(3) Press TI ENTRY MOD FREQ pushbutton and enter DATA modulated
frequency for each row in table 22.  Set audio analyzer to measure distortion.  Audio
analyzer will indicate within limits listed in table 22.