SECTION IV. CALIBRATION PROCESS FOR FLUKE 80K-6 HIGH VOLTAGE PROBE

Calibration Procedure For Digital Multimeters An/Psm-45A (27/Fm) And Models 27 And 27/An And High Voltage Probe, Type 80K-6 Page Navigation
  6    7    8    9    10  11
img
. . . .
TB 9-6625-2190-24
14. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S)  to  minimum  after  each  step  within  the
performance checks where applicable.
WARNING
Always hold the TI by its black handle.  When making a
measurement, do not make body contact with the TI tip or the
red portion of TI. TI ground clip lead must be attached to earth
ground.
NOTE
To calibrate the high voltage probe when not part of a set,
customer must supply a handheld multimeter with input
impedance of 10 MΩ 10%.
15. Dc Voltage
a. Performance Check
(1) Connect TI cable to the handheld meter VΩ
and COM observing correct polarity.
(2) Set handheld meter to 3 V range.
(3) Connect the TI ground clip lead to calibrator OUTPUT LO. Hold the probe by
the black handle and connect the probe tip to calibrator OUTPUT HI.
(4) Set calibrator to 1000 V.
(5) Push calibrator OPR/STBY button to OPR, handheld multimeter should
indicate between .990 and 1.010 V. No calibration adjustments can be made on the probe.
(6) Set calibrator to STBY.
b. Adjustments. No adjustments can be made.
16. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form with TB 750-25.
11/(12 Blank)