TB 9-6625-2190-24
Table 5. Ac Voltage AN/PSM-45A (FLUKE 27/FM) - Continued
32.00
V
30
V
100 Hz
29.80
V
30.20
V
32.00
V
30
V
900
Hz
29.80
V
30.20
V
32.00
V
30
V
5
kHz
28.45
V
31.55
V
320.0
V
300
V
40
Hz
295.0
V
305.0
V
320.0
V
300
V
100
Hz
298.0
V
302.0
V
320.0
V
300
V
900
Hz
298.0
V
302.0
V
320.0
V
300
V
5
kHz
284.5
V
315.5
V
1000
V
1000
V
40
Hz
980
V
1020
V
1000
V
1000
V
100
Hz
990
V
1010
V
1000
V
1000
V
900
Hz
990
V
1010
V
1000
V
1000
V
5
kHz
945
V
1055
V
1
V and press RANGE pushbutton for remaining range settings.
Set function switch to
Calibrator settings
Test instrument indications
Test instrument
range settings
Voltage
Min
Max
320.0
mV
300
mV
40
Hz
298.2
mV
301.8
mV
320.0
mV
300
mV
1.9 KHz
298.2
mV
301.8
mV
320.0
mV
300
mV
9.9 KHz
293.7
mV
306.3
mV
320.0
mV
300
mV
30
KHz
287.0
mV
313.0
mV
3.200 V1
3
V
40
Hz
2.982 V
3.018
V
3.200 V
3
V
1.9 KHz
2.982 V
3.018
V
3.200 V
3
V
9.9 KHz
2.937 V
3.063
V
3.200 V
3
V
30
KHz
2.870 V
3.130
V
32.00
V
30
V
40
Hz
29.82
V
30.18
V
32.00
V
30
V
1.9 KHz
29.82
V
30.18
V
32.00
V
30
V
9.9 KHz
29.37
V
30.63
V
32.00
V
30
V
30
KHz
28.70
V
31.30
V
320.0
V
300
V
40
Hz
298.2
V
301.8
V
320.0
V
300
V
1.9 KHz
298.2
V
301.8
V
320.0
V
300
V
9.9 KHz
293.7
V
306.3
V
320.0
V
300
V
30
KHz
287.0
V
313.0
V
1000
V
1000
V
40
Hz
987
V
1013
V
1000
V
1000
V
1.9 KHz
987
V
1013
V
1000
V
1000
V
9.9 KHz
967
V
1033
V
1
V and press RANGE pushbutton for remaining range settings.
Set function switch to
b. Adjustments
V.
(1) Set function switch to
(2) Press RANGE pushbutton for 3.200 V range.
(3) Set calibrator for a 3 V, 100 Hz output.
8