SECTION III. CALIBRATION PROCESS FOR MULTIMETERS
Figure 1. Adjustment locations.

Calibration Procedure For Digital Multimeters An/Psm-45A (27/Fm) And Models 27 And 27/An And High Voltage Probe, Type 80K-6 Page Navigation
  1    2    3    4  5  6    7    8    9    10  
img
. . . . . . .
TB 9-6625-2190-24
7. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S)  to  minimum  after  each  step  within  the
performance checks where applicable.
a. Remove protective cover from TI as necessary to gain access to adjustments.
b. For some checks throughout this procedure, it may be necessary to manually set TI
range by pressing RANGE pushbutton.
8. Dc Voltage
a. Performance Check
(1) Connect calibrator OUTPUT HI and LO to TI VΩ
and COM, observing
correct polarity.
mV. Set calibrator to first setting in table 3. Push
(2) Set function switch to
calibrator OPR/STBY button to OPR.
V. Set TI and calibrator to the remaining settings in
(3) Set TI function switch to
table 3.
(4) Set calibrator to STBY. If TI did not indicate within limits specified, in table 3,
perform b below before proceeding.
Table 3. Dc Voltage
Test instrument indications
Test instrument
range settings
Calibrator settings
Min
Max
320.0
mV
300
mV
299.6
mV
300.4
mV
3.200  V
3
V
2.996 V
3.004 V
32.00  V
30
V
29.96  V
30.04  V
320.0
V
300
V
299.6
V
300.4
V
1000
V
1000
V
998
V
1002
V
NOTE
Model 80K-6 high voltage probe is a component part of
the AN/PSM-45A (FLUKE 27/FM) meter set.
(5) If high voltage probe (80K-6) was supplied continue with step (6) below, if not
furnished with meter proceed to step (11).
(6) Disconnect connections made in step (1).
CHANGE 1 5