TB 9-6625-2190-24
Test instrument parameters
Performance specifications
Dc voltage
Range:
0 to 1000 V in 5 ranges (5000 V using high voltage probe)
(0.1% of reading + 1 digit)
Accuracy:
(1% of reading using high voltage probe)
Dc current
Range:
0 to 10 A in 5 ranges
(0.75% of reading + 2 digits)1
Accuracy:
Ac voltage
AN/PSM-45A (FLUKE 27/FM)
Range:
mV to 1000 V in 5 ranges
(1.5% of reading + 5 digits)
Accuracy: 20 Hz to 40 Hz;
(0.5% of reading + 5 digits)
40 Hz to 1 KHz;
(5% of reading + 5 digits)
1 KHz to 5 KHz;
27 and 27/AN
Range:
mV to 320 V
Accuracy: 40 Hz to 2 KHz
(0.5% of reading + 3 digits)
2 KHz to 10 KHz
(2% of reading + 3 digits)
10 KHz to 30KHz
(4% of reading + 10 digits)
Range:
1000 V
Accuracy:
40 Hz to 2 KHz
(1% of reading + 3 digits)
2 KHz to 10 KHz
(3% of reading + 3 digits)
Ac current2
Range:
0 to 10 A in 5 ranges
Accuracy:
AN/PSM-45A (FLUKE 27/FM)
(1.5% of reading + 5 digits)
0 - 320mA
40 Hz to 1 kHz
10A
40 Hz to 1 KHz
(2.5% of reading + 5 digits)
Accuracy:
27 & 27/AN
All ranges
40 Hz to 1 KHz
(1.5% of reading + 2 digits)
0 Ω to 32 MΩ in 6 ranges
Range:
Accuracy: 320 Ω range
(0.3% of reading + 2 digits)
AN/PSM-45A (FLUKE 27/FM)
3.2 KΩ to 3.2 MΩ ranges
(0.25% of reading + 1 digit)
32 MΩ range
(1% of reading + 1 digit)
Accuracy: 320 Ω range
27 & 27/AN
(0.3% of reading + 3 digits)
3.2 KΩ to 3.2 MΩ ranges
(0.2% of reading + 1 digit)
32 MΩ range
(1% of reading + 1 digit)
1
Specifications for AN/PSM-45A (FLUKE 27/FM) based on Army Specification procurement document.
2
Ac current verified by dc current check, current measurements of ac and dc are made using same internal shunt resistor.
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286; AN/GSM-287; or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily