Figure 4. Input amplifier - (Hewlett-Packard, Model 5335A) adjustment waveform.
Figure 5. Test instrument - (Hewlett-Packard, Model 5335A) DVM right side view

Calibration Procedure For Universal Frequency Counter, Model 5335A Options 010, 020, 030, And 040 Page Navigation
  3    4    5    6    7  8  9    10    11  
TB 9-6625-2214-24
(2) Press FUNCTION pushbutton to FREQ C.
(3) Adjust signal generator frequency to 150 MHz and slowly increase output until
TI displays a stable indication of applied frequency. If signal generator output exceeds 10
mV, perform b below.
(4) Repeat technique of (3) above, varying signal generator frequency from 150 MHz
to 1 GHz.
(5) Adjust signal generator frequency to 1 GHz and slowly increase output until TI
displays a stable indication of applied frequency.  If signal generator output exceeds
100mV, perform b below.
(6) Repeat technique of (5) above, varying signal generator frequency from 1 GHz to
1.3 GHz.
b. Adjustments. Adjust signal generator frequency to 1 GHz and output to 50 mV.
Adjust A9 R22 (A9 board lower center, top right of TI) until TI displays a stable indication
of applied frequency (R).
11. Digital Voltmeter (Option 020)
a. Performance Check
(1) Set TI controls as listed in (a) through (g) below:
FUNCTION O/GATE TIME to GATE TIME.
(a)
MATH to OFF (LEDs off).
(b)
STATISTICS to OFF (LEDs off).
(c)
GATE MODE/ADJ/NORM/FAST to NORM.
(d)
CYCLE to NORM.
(e)
(f)
Adjust gate time to 150 ms.
FUNCTION pushbutton to VOLTS.
(g)
(2) Connect short between DCV and COM connector. TI will display 0 V 2 mV. If
not perform b below.
(3) Remove short and connect calibrator to DCV and COM.  Set calibrator for
4.0000 V. TI display will indicate between 3.996 and 4.004 V dc. Set calibrator for -4.0000
V. TI display will indicate between -3.996 and -4.004 V dc. If not perform b below.
(4) Set calibrator for 40.000 V. TI display will indicate between 39.96 and 40.04 V
dc. Set calibrator for -40.000 V. TI display will indicate between -39.96 and -40.04 V dc. If
not perform b below.
(5) Set calibrator for 400.00 V. TI display will indicate between 399.6 and 400.4 V
dc. Set calibrator for -400.00 V. TI display will indicate between -399.6 and -400.4 V dc. If
not perform b below.