Table 1. Calibration Description.
SECTION III. CALIBRATION PROCESS FOR MEASURING SYSTEM, TD-1209/U (HP 5300A) AND 50 MHZ UNIVERSAL COUNTER, TD-1211/U (HP 5302A)

Calibration Procedure For Measuring System, Td-1209/U (Model 5300A); 50 Mhz Universal Counter, Td-1211/U (Model 5302A); 10 Mhz Counter (Model 5301A); 500 Mhz Counter (Model 5303A); 525 Mhz Counter (Model 5303B); Timer/Counter (Model 5304A) And Multimeter/Counter (Model 5306A) Page Navigation
  1    2    3    4  5  6    7    8    9    10  
img
. . . . . . . .
TB 9-6625-2215-24
Table 1. Calibration Description Continued.
Test instrument
Performance specifications
parameters
Hewlett-Packard, Model 5304A
Frequency
Range:
0 to 10 MHz (dc coupled)
100 Hz to 10 MHz (ac coupled)
Sensitivity
25 mV rms:
0 to 1 MHz
50 mV rms:
1 to 10 MHz
Hewlett-Packard, Model 5306A
Dc voltage
Range:
0 to 1000 V in 3 ranges
Accuracy:
10 and 100 V ranges:
(0.03% of reading +0.003% of range)
1000 V range:
(0.097% of reading
+0.03% of range)
Ac voltage
Range:
0 to 1000 V in 3 ranges
Accuracy:
10 V range:
40 Hz to 10 kHz (0.98% of reading
+0.02% of range);
10 to 100 kHz (0.98% of reading +0.02% of range)
100 and 1000 V range:
40 to 500 Hz (1.5% of reading +0.05% of range)
Resistance
Range:
10 kΩ, 100 kΩ, and 10 MΩ,
Accuracy:
10 and 100 kΩ (0.5% of reading +0.003% of range)
10 MΩ range (0.75% of reading +0.003% of range)
1Not
calibrated above 500 MHz.
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure.  This equipment is issued with Secondary Transfer Calibration
Standards Set, AN/GSM-286; AN/GSM-287; or AN/GSM-440. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI.  Where the four-to-one ratio
cannot be met, the actual accuracy of the equipment selected is shown in parentheses.
5. Accessories Required.  The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure.