Table 3. Sensitivity OUTPUT A 50 MHz.
SECTION IV. CALIBRATION PROCESS FOR 10 MHZ COUNTER (HP 5301A)

Calibration Procedure For Measuring System, Td-1209/U (Model 5300A); 50 Mhz Universal Counter, Td-1211/U (Model 5302A); 10 Mhz Counter (Model 5301A); 500 Mhz Counter (Model 5303A); 525 Mhz Counter (Model 5303B); Timer/Counter (Model 5304A) And Multimeter/Counter (Model 5306A) Page Navigation
  4    5    6    7    8  9  10    11    12    13    14  
img
. . . . .
TB 9-6625-2215-24
(10) Repeat technique of (8) and (9) above for function generator frequencies listed in
table 4. Function generator adjusted output will be within limit specified in table 4.
Table 4. Sensitivity OUTPUT B 10 MHz.
Function generator output
Frequency
Amplitude limits
(Hz)
(mV)
≤ 50
10
≤ 25
50
≤ 25
100
≤ 25
1000
≤ 25
10000
≤ 25
100000
≤ 25
1000000
≤ 50
100000001
Set TI TIME BASE switch to AUTO.
1
b. Adjustments
NOTE
For serial number prefix below 1444A, no adjustments can be
made. For serial number prefix 1444A and above, perform (1)
through (4) below.
(1) Remove bottom cover from TI.
(2) Adjust function generator for 100 kHz and amplitude of 25 mV.
(3) Adjust A1 R59 (A channel) closest to front panel or A1 R57 (B channel) near rear
of TI to obtain stable indication of applied frequency.
(4) While reducing function generator amplitude, adjust A1 R59 (A channel) or A1 R57
(B channel) to obtain maximum sensitivity.
10. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
8