TB 9-6625-2215-24
(c) A SENSITIVITY control fully cw.
(d) TIME BASE switch to 10S.
(3) Adjust function generator for a 10 Hz sine wave frequency signal and amplitude
to minimum output.
(4) Slowly increase function generator amplitude until TI measuring system
indicates a stable count of the applied frequency. Function generator adjusted output will
be within limit specified in first row of table 3.
(5) Repeat technique of (3) and (4) above for function generator frequencies listed in
table 3. Function generator adjusted output will be within limit specified in table 3.
Function generator output
Frequency
Amplitude limits
(Hz)
(mV)
≤ 50
10
≤ 25
50
≤ 25
100
≤ 25
1000
≤ 25
10000
≤ 25
100000
≤ 25
1000000
≤ 50
100000001
≤ 100
50000000
Set TI TIME BASE switch to AUTO.
1
(6) Set function generator output to minimum and connect to TI B 10 MHz jack,
using termination.
(7) Position TI 50 MHz counter controls as listed in (a) through (d) below:
(a) FUNCTION switch to FREQ B.
(b) B SENSITIVITY control fully cw.
(d) TIME BASE switch to 10S.
(8) Adjust function generator for a 10 Hz sine wave frequency signal and amplitude
to minimum output.
(9) Slowly increase function generator amplitude until TI measuring system
indicates a stable count of the applied frequency. Function generator adjusted output will
be within limit specified in first row of table 4.
7