TB 9-6625-2215-24
(2) Connect function generator output to TI INPUT, using termination.
to minimum output.
(4) Slowly increase function generator amplitude until measuring system indicates
a stable count of the applied frequency. Function generator output will be within limit
(5) Repeat technique of (3) and (4) above for function generator frequencies listed in
table 5. Function generator adjusted output will be within limit specified in table 5.
Function generator output
Frequency
Amplitude limits
(Hz)
(mV)
≤ 50
10
≤ 25
50
≤ 25
100
≤ 25
1000
≤ 25
10000
≤ 25
1
100000
≤ 25
10000002
≤ 50
3
10000000
Set TI GATE switch to 1S.
1
Set TI GATE switch to .1S.
2
3 Set TI GATE switch to AUTO.
b. Adjustments. No adjustments can be made.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.