SECTION IV. CALIBRATION PROCESS FOR 10 MHZ COUNTER (HP 5301A)
SECTION V. CALIBRATION PROCESS FOR 500 MHZ COUNTER (HP 5303A)

Calibration Procedure For Measuring System, Td-1209/U (Model 5300A); 50 Mhz Universal Counter, Td-1211/U (Model 5302A); 10 Mhz Counter (Model 5301A); 500 Mhz Counter (Model 5303A); 525 Mhz Counter (Model 5303B); Timer/Counter (Model 5304A) And Multimeter/Counter (Model 5306A) Page Navigation
  6    7    8    9    10  11  12    13    14    15    16  
img
. . . . . .
TB 9-6625-2215-24
(2) Connect function generator output to TI INPUT, using termination.
(3) Adjust function generator for a 10 Hz sine wave frequency signal and amplitude
to minimum output.
(4) Slowly increase function generator amplitude until measuring system indicates
a stable count of the applied frequency. Function generator output will be within limit
specified in first row of table 5.
(5) Repeat technique of (3) and (4) above for function generator frequencies listed in
table 5. Function generator adjusted output will be within limit specified in table 5.
Table 5. Sensitivity.
Function generator output
Frequency
Amplitude limits
(Hz)
(mV)
≤ 50
10
≤ 25
50
≤ 25
100
≤ 25
1000
≤ 25
10000
≤ 25
1
100000
≤ 25
10000002
≤ 50
3
10000000
Set TI GATE switch to 1S.
1
Set TI GATE switch to .1S.
2
3 Set TI GATE switch to AUTO.
b. Adjustments. No adjustments can be made.
14. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.