SECTION II. EQUIPMENT REQUIREMENTS
Figure 1. Access and adjustments.

Calibration Procedure For Digital Multimeter Model 8060A Page Navigation
  1    2  3  4    5    6    7  
img
. . . . . . . . .
TB 9-6625-2230-24
SECTION III
CALIBRATION PROCESS
6. Preliminary Instructions
a. The instructions outlined in paragraph 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in table 2.
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure.  Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
7. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S)  to  minimum  after  each  step  within  the
performance check where applicable.
a. Set ON-OFF switch to ON.
b. Set TI to measure dc voltage.
c. Press 200 mV pushbutton.
8. Dc Voltage
a. Performance Check
(1) Connect calibrator output terminals to TI VΩS and COMMON terminals.
(2) Press TI range pushbutton settings and set calibrator output as specified in
table 3. TI will indicate within the specified limits.
Table 3. Dc Voltage
Test instrument
Calibrator
range
Test instrument indications
Adjustments
output
pushbutton
(fig. 1)
settings
settings
Min
Max
200 mV
190 mV
189.91 mV
190.09 mV
b(1)
4