Table 3. Dc Voltage - Continued
Test instrument
Calibrator
Test instrument indications
Adjustments
range
output
pushbutton
settings
settings
2V
1.90
V
1.8991 V
1.9009 V
b(2) & (3)
20 V
19.000 V
18.989
V
19.011
V
200 V
190.00
V
189.89
V
190.11
V
1000 V
900
V
899.4
V
900.6
V
b. Adjustments
(1) Set calibrator to 190 mV. Press TI 200 mV pushbutton and adjust R8 (fig. 1) for
a TI indication of 190.00 (R).
(2) Press 2V pushbutton. Adjust R6 (fig. 1) fully cw and adjust R5 (fig. 1) ccw.
(3) Set calibrator output to 1.9000 V. Adjust R5 (fig. 1) for a TI indication slightly
greater than 1.9000. Adjust R6 (fig. 1) for a TI indication of 1.9000 (R).