SECTION III. CALIBRATION PROCESS
Figure 1. Dc module adjustment.

Calibration Procedure For Radio Test Set Ts-24(B) Page Navigation
  1    2    3  4  5    6    7    8    9  
img
. . . . . . . .
TB 9-6625-2237-24
a. Set the measuring receiver to measure power. Connect sensor module to signal
generator RF OUTPUT.
b. Adjust signal generator frequency controls to signal generator frequency settings
listed in table 4.
c. Adjust  signal  generator  amplitude  controls  for  a  measuring  receiver
FREQUENCY/LEVEL display indication equal to the value listed in table 4 test settings.
d. Record signal generator indication in signal generator amplitude output indication
column of table 4.
e. Repeat b through d above for remaining test settings listed in table 4.
f. Disconnect measuring receiver sensor module from signal generator RF OUTPUT.
Table 4. Signal Generator Characterization
Signal generator
Amplitude output
indications
Frequency
Test settings
settings
(dBm)
(dBm)
(MHz)
19.1
243.0
19.6
243.0
18.6
243.0
19.1
282.8
19.6
282.8
18.6
282.8
15.0
282.8
8. Dc Module Check
a. Performance Check
(1) Close and latch TI test chamber cover. Remove any batteries from PRC-90/106
battery chamber.
(2) Connect dc power supply to EXT. PWR PRC-90/106 (fig. 1) test points.
NOTE
Monitor dc power supply with multimeter (dc mode).
(3) Connect calibration plug to TI connector 2J1 on right side of dc module.
(4) Adjust dc power supply for a 10.90 V dc indication on multimeter.
(5) The TI BATTERY VOLTAGE PASS indicator must be illuminated and the
FAIL indicator must be off; if not perform b (1) below.
(6) Adjust dc power supply for a 10.70 V dc indication on multimeter.